Detail EUTECH CON450 CONDUCTIVITY METER
Features:
Conductivity / Total Dissolved Solids / Salinity / Temperature
Backlit, rugged waterproof meter is ideal for any environment
- Durable waterproof and dustproof design with IP67 rating
- Large, backlit custom display – View measurement with temperature and cell efficiency
- Grip-Clip™ – Quickly and easily secures the probe to your beaker or container
- Calibrate using automatically recognized standards or manually; to a single point across all ranges or to multiple points ( one calibration point per each conductivity range)
- Selectable cell constant – 0.10, 1.00, or 10.0
- Adjustable temperature coefficient from 0.00 to 10.00 % per º C. Set to 0.0 % to meet USP methods for non-compensated measurements
- Adjustable conductivity-to-TDS factor – Adjustable 0.4 to 1.0
- Selectable º C / º F – Choose units to best fit your application
- Includes our easy to clean conductivity cell
- Ready indicator, manual or automatic measurement hold feature and optional automatic shutoff function
- Store up to 500 data sets with data – Use USB or RS232 output with optional data cable to send data to your software or our free communication software
- Built-in real-time clock stamps stored data and calibration data with time and date – Meets standards for GLP ( Good Laboratory Practices)
- Built-in stand with wall-mount – Allows for easy benchtop measurement or storage
Specifications:
Conductivity
Range 0 to 200 mS
Resolution 0.01 µ S to 0.1 mS
Accuracy ± 1 % full scale
Calibration
Auto/ manual; 1 point/ range
TDS
Range 0 to 200 ppt ( TDS factor 1.0)
Resolution 0.01 ppm to 0.1 ppt
Accuracy ± 1 % full scale
Calibration Manual; 1 point/ range
Salinity
Range 0 to 50 ppt
Resolution 0.01
Accuracy
± 1 % full scale
Calibration Manual; 1 point
Temperature
Range -10.0 to 110.0 º C ( 14.0 to 230.0 º F)
Resolution 0.1 º C/ º F
Accuracy ± 0.5 º C; ± 0.9 º F
Calibration Offset in 0.1 increments, ± 10 º C or ± 18 º F
Others
Display Custom LCD with backlight
Memory Stores up to 500 data sets
Output USB or RS232
Real-Time Clock Time and date stamp on calibration and stored data
Temperature Compensation Automatic or manual from -10 to 110 º C
Power Two AA batteries ( included) , > 300 hours ( without backlight) ; AC adapter ( optional)
Dimensions ( L x W x H)
Meter: 7 7/ 8 " x 3 1/ 8 " x 2 1/ 4 " ( 20 x 8.3 x 5.7 cm)
Boxed: 9 1/ 8 " x 9 1/ 8 " x 2 3/ 4 " ( 23 x 23 x 7 cm)
Weight Meter: 1 lb ( 0.5 kg)
Boxed: 1.4 lb ( 0.7 kg)