Detail HORIBA OCMA350 OIL CONTENT ANALYZER
Just by setting the equipment, the OCMA-350 provides highly accurate analysis of the oil content of water samples from which the oil has already been extracted in the auxiliary cell. In addition to displaying residual oil and solid body content suspended in water in a range from 0mg/ l-200mg/ l, the OCMA-350 allows users to display analysis results for land in mg/ kg and for oil absorbency in Abs. The OCMA-350 is also well suited for evaluating the capabilities of semiconductor fabrication and precision machinery cleaning equipment, evaluating the properties of industrial process oil and the residual oil of polishing materials as well as wastewater that has been adulterated with silt, sludge, and other suspended particles.
Specifications
Measurement principle: Solvent extraction, NDIR
analysis ( Infrared spectroscopy)
Measurement range: 0 to 200 mg/ L, 0 to 1000 mg/ kg, 0 to 1 Abs.
Resolution mg/ L;
0 to 99.9 mg/ L; 0.1 mg/ L
100 to 200 mg/ L; 1 mg/ L
mg/ kg;
0 to 9.99 mg/ kg; 0.01 mg/ kg
10.0 to 99.9 mg/ kg; 0.1 mg/ kg
100 to 1000 mg/ kg; 1 mg/ kg
Abs;
0 to 1000 Abs; 0.001 Abs
Repeatability mg/ L;
0 to 9.9 mg/ L; ± 0.4 mg/ L ± 1 digit
10.0 to 99.9 mg/ L; ± 2.0 mg/ L ± 1 digit
100 to 200 mg/ L; ± 4 mg/ L ± 1 digit
Abs;
± 1% of full scale
Measurement Manually controlled
Calibration Automatic calibration after the calibration standard is introduced to the instrument
Extraction Solvent S-316 solvent
Sample/ Solvent Volume Approx. 6.5 mL, ratio of sample to solvent is 1: 1
Extraction Method Built-in extractor for oily water samples
Display Measured value; 3 1/ 2digits LCD with back-light
Message; Character display LCD with back-light ( 16 x 2 characters)
Functions Self diagnostics, Auto hold function, Calendar clock
Output RS-232C and centronics printer port
Ambient Temperature 0 to 40
Power 100 to 240 V AC ± 10% , 50/ 60 Hz, 60 VA
Dimensions 200 ( H) x 250 ( W) x 285 ( D) mm
Mass Approx. 5 kg
Tampilkan Lebih Banyak